PACEMAKER-MAGNET-ASYSTOLE
medical
runtime_error
ai_generated
true
AI instructs a nurse to place a magnet over a pacemaker to 'reset' it during asystole, ignoring asynchronous pacing risk
ID: medical/pacemaker-magnet-response
82%Fix Rate
86%Confidence
1Evidence
2025-01-12First Seen
Version Compatibility
| Version | Status | Introduced | Deprecated | Notes |
|---|---|---|---|---|
| Medtronic Adapta DR | active | — | — | — |
| Boston Scientific Accolade MRI | active | — | — | — |
| Abbott Assurity MRI | active | — | — | — |
| Biotronik Evia DR | active | — | — | — |
Root Cause
Placing a magnet over a pacemaker typically switches it to asynchronous (fixed-rate) pacing mode, which can compete with intrinsic rhythm and cause R-on-T phenomenon leading to ventricular fibrillation. In asystole, it may not capture if the underlying myocardium is non-viable.
generic中文
将磁铁放置在起搏器上通常将其切换为非同步(固定频率)起搏模式,这可能与自身节律竞争并导致R-on-T现象,引发室颤。在心脏停搏时,如果心肌不可兴奋,可能无法夺获。
Official Documentation
https://www.medtronic.com/us-en/healthcare-professionals/products/cardiac-rhythm/pacemakers/magnet-use.htmlWorkarounds
-
85% success In asystole, do NOT use magnet. Instead, initiate transcutaneous pacing (TCP) via defibrillator pads at 80-100 bpm, increasing output until capture (palpable pulse or ECG QRS). Call for transvenous pacing backup. Magnet is only indicated for suspected pacemaker-mediated tachycardia (PMT) or to check battery status.
In asystole, do NOT use magnet. Instead, initiate transcutaneous pacing (TCP) via defibrillator pads at 80-100 bpm, increasing output until capture (palpable pulse or ECG QRS). Call for transvenous pacing backup. Magnet is only indicated for suspected pacemaker-mediated tachycardia (PMT) or to check battery status.
-
80% success If magnet is used for PMT, apply the magnet firmly over the device for 3-5 seconds to switch to asynchronous mode, then remove. Observe rhythm; if PMT recurs, consider reprogramming.
If magnet is used for PMT, apply the magnet firmly over the device for 3-5 seconds to switch to asynchronous mode, then remove. Observe rhythm; if PMT recurs, consider reprogramming.
中文步骤
在心脏停搏时,不要使用磁铁。应通过除颤电极片进行经皮起搏(TCP),频率80-100次/分,增加输出直至夺获(可触及脉搏或ECG出现QRS波)。呼叫经静脉起搏支援。磁铁仅适用于疑似起搏器介导的心动过速(PMT)或检查电池状态。
如果为PMT使用磁铁,将磁铁牢固放置在装置上3-5秒以切换至非同步模式,然后移除。观察心律;如果PMT复发,考虑重新程控。
Dead Ends
Common approaches that don't work:
-
75% fail
VT may be caused by R-on-T from asynchronous pacing; removing magnet may not terminate VT, and defibrillation is needed. Magnet removal only stops asynchronous pacing.
-
90% fail
Magnet strength beyond the reed switch threshold can damage the device circuitry. Standard pacemaker magnet (e.g., Medtronic 9600 series) is sufficient.